Respironics 920M Instrukcja Operatora

Przeglądaj online lub pobierz Instrukcja Operatora dla Pomiary, testowanie i kontrola Respironics 920M. Respironics 920M Operator`s manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Operator’s Manual

Operator’s ManualHandheld Pulse Oximeter920M TMPLUS1English/Español

Strona 2

4 General DescriptionGeneral DescriptionThe 920M PLUS is a digital handheld pulse oximeter that displays numerical values for blood oxygen saturation

Strona 3

General Description 5Because the 920M PLUS Handheld Pulse Oximeter has no patient alarms, frequent observation of SpO2 and pulse rate displays is requ

Strona 4 - Guide to Symbols

6 Basic OperationBasic OperationInstalling and Using the BatteriesThe 920M PLUS Handheld Pulse Oximeter can be powered by 4 AA-size alkaline batteries

Strona 5 - Table of Contents

Basic Operation 7Important Notes about Battery UseFour AA alkaline batteries provide the 920M PLUS with approximately 100 hours of continuous operatio

Strona 6

8 Basic OperationNote:After removing batteries from the 920M PLUS, you will haveapproximately 2 minutes to replace them before losing data stored inme

Strona 7 - Precautions for Use

Basic Operation 9Power On/StandbyTurn on the 920M PLUS by pressing and releasing the On/Standby switch . Turn off the unit by pressing and holding thi

Strona 8 - 2 Precautions for Use

10 Basic OperationMonitoringVerify that the pulse oximeter sensor is properly positioned (on a finger, for example). Ensure that the pulse oximeter is

Strona 9 - Indications for Use

Detailed Operation 11Detailed OperationSwitches and FunctionsAll functions of the 920M PLUS are controlled by the On/Standby and advance switches f

Strona 10 - 920M PLUS

12 Detailed OperationMaking Selections in Setup Mode1. Upon entering setup mode, the displays will show . (Memory clear and the default value “no”, t

Strona 11 - KENNESAW, GA USA

Detailed Operation 13Clearing MemoryUpon entering setup mode, “Clr” will appear in the SpO2 display, indicating the memory clear function. (Refer to “

Strona 13 - Basic Operation 7

14 Detailed OperationCalendar and Clock SettingsNote:Setting the month to “00” disables the calendar and clock functions and helps conserve battery li

Strona 14 - Connecting the Sensor

Detailed Operation 15Displays and IndicatorsSpO2 DisplayThe SpO2 display is the upper numeric display. (Identified by the %SpO2 symbol; see Figure 2 o

Strona 15 - Power On/Standby

16 Detailed OperationLow Battery IndicatorWhen the batteries are low, the low battery indicator, a yellow LED, will be steadily illuminated. (Identifi

Strona 16 - Monitoring

Communications 17CommunicationsSerial OutputThe 920M PLUS provides real-time data output capability via the pulse oximeter sensor connector (a 9-pin S

Strona 17 - Detailed Operation

18 CommunicationsEach time the 920M PLUS is turned on, the current time/date information (if the clock is set properly) is stored in memory to allow q

Strona 18

Specifications 19Specifications920M PLUS SpecificationsOxygen Saturation Range (SpO2) 0 to 100%Pulse Rate Range 18 to 300 Pulses Per MinuteAccuracySpO

Strona 19 - (default)

20 Service and SupportService and SupportCAUTION: The 920M PLUS is a precision electronic instrument andmust be repaired by trained Respironics person

Strona 20 - Confirm clearing the memory?

Warranty 21WarrantyRESPIRONICS, INC., (Respironics) warrants to the purchaser, for a period of three years from the date of purchase, each Model 920M

Strona 21 - Displays and Indicators

22 AccessoriesAccessoriesThe following Respironics accessories function with the 920M PLUS:Part Number Description1007355 Rechargeable NiMH Battery Pa

Strona 22 - Sensor Fault Indication

Accessories 23For more information about Respironics parts and accessories, contact your distributor, or contact Respironics at (800) 345-6443 (USA an

Strona 23 - Communications

The information in this manual has been carefully checked and is believed to be accurate. In the interest of ongoing product development, Respironics

Strona 24 - Memory Playback

24 Troubleshooting GuideTroubleshooting GuideTroubleshooting GuideProblem Possible Cause Possible SolutionThe unit will not turn on. The batteries are

Strona 25 - Specifications

Troubleshooting Guide 25You are unable to obtain a green blinking pulse quality display.Low patient pulse strength;orthe sensor site is poorly perfuse

Strona 26 - Service and Support

Guide to SymbolsFront Panel SymbolsPercent Oxygen SaturationPulse RatePulse Quality (tricolor LED)Low BatteryFront Panel SwitchesOn/StandbyAdvanceOthe

Strona 27 - Warranty

Table of ContentsPrecautions for Use ...1Indications for Use ...

Strona 29 - (724) 387-4000

Precautions for Use 1Precautions for UseContraindications•Do not use the 920M PLUS in an MRI environment.• Do not use the 920M PLUS in situations wher

Strona 30 - Troubleshooting Guide

2 Precautions for Use• The 920M PLUS Pulse Oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that not

Strona 31 - 3485-000-03

Indications for Use 3Indications for UseThe Respironics® 920M™ PLUS Handheld Pulse Oximeter is indicated for measuring and displaying functional oxyge

Komentarze do niniejszej Instrukcji

Brak uwag