15
Power On Self-Test
When the Model 920M PLUS is turned on for normal operation, the unit
will cycle through a startup/initialization sequence before displaying valid
data. During startup, always check for any missing indicators or LED
display segments. If any indicator is not functioning, do not use the device.
Contact Respironics’ Technical Service for repair or replacement.
During its normal startup sequence, the device will cycle as follows:
•“888 888” appears briefly in the SpO
2
and Pulse Rate displays.
• the amber Low Battery LED turns on steadily for a few seconds.
• the Pulse Quality display turns red for 1 second, then green for
1 second, then shuts off.
• the clock time currently set in the memory (in hours and minutes,
04 41 for example) appears briefly in the displays.
• the software revision numbers (display in the following order, each for
approximately 1 second): Main revision “r” + 3 digit; Memory revision
“n” “n” (for m) + 3 digits.
• (two dashes) appear in the displays until a valid pulse signal is
detected.
NOTE: This startup sequence varies slightly when entering setup mode
at power on.