Memory Download
NOTE: Downloading the data in memory does not clear the memory.
Downloading the Data Stored in Memory
1. With the unit off, press and hold the Advance button while
pressing and then releasing the On/Off button .
2. Release the advance button when 888 888 is displayed on the
SpO
2
and pulse rate displays. The clock time currently set in the
memory (04 41 for example) appears briefly in the displays, and
then CLr no appears.
3. Data will be automatically downloaded from memory. Data is
downloaded at a rate of 20 minutes of collected data per second. A
72-hour recording session (the maximum memory saved) is
downloaded in approximately 3.5 minutes.
4. After downloading is complete, the device should be shut off before
collecting new patient data.
5. The patient data is held in memory as long as the batteries are
sufficiently charged. (See Clearing the Memory section.) To clear
memory, use the memory clear function.